Pipette Common Problem Solving Method

The equipment and equipment in the laboratory are as comprehensive as the natural world. They say that the sparrows are small and complete, but in the laboratory, the same small but indispensable instrument has a micropipette for biochemistry. An essential tool for experimental or molecular biology experiments.

The pipette is a precision liquid-removing device that can be adjusted freely within a certain capacity range. It is commonly called a pipetting gun, but there are some common problems such as dripping, leaking, and inaccurate during use.

1. Dropping Liquid (Dripping)

Cause: Worn out or damaged seal and O-rings, improper pipette assembly, or using the wrong tips.

Solution:
Inspect and replace the seal and O-rings if they are worn or damaged.
Ensure the pipette is properly assembled according to the manufacturer’s instructions.
Use tips recommended by the pipette manufacturer.

There are many possible reasons for “dropping”, usually due to the removal of special liquids. Some liquids with high viscosity, high density, and high saturated vapor pressure are often the “culprits” of dripping. When removing such liquids, we recommend a reverse pipetting method.

2. Leakage

Cause: Improper tip attachment, damaged tip, or damaged piston.

Solution:
Attach the tip firmly and ensure it is seated properly.
Check the tip for cracks or defects and replace them if necessary.
Inspect the piston for damage and replace it if required.

The main reason for “leakage” is that the pipette is poorly sealed and the tips are not matched. If the experiment repeatedly leaks, the leakiness of the pipette itself is eliminated, and the matching of the tips needs to be considered. Usually, the same brand of pipettes and tips are matched to achieve a good system seal.

3. Inaccurate Volume Dispensing (Unacceptable Performance)

Cause: Calibration issues, user technique, or environmental factors.

Solution:
Regularly calibrate the pipette following the manufacturer’s guidelines.
Train users on proper pipetting techniques, including pre-wetting tips, consistent plunger speed, and proper angle.
Ensure the pipette is used in a controlled environment, avoiding temperature fluctuations and vibrations.

The main reason for “unacceptable” is that the pipette is not kept vertical when aspirating, or the absorbed liquid is too thick and volatile. Your pipette may need to be calibrated when you feel the aspiration is inaccurate during the experiment. ISO 9001 and GLP require periodic pipette calibration (approximately every 3-12 months), and pipetting products should be calibrated using the gravimetric method by ISO 8655. In daily use, keeping the pipette vertical when aspirating is also necessary. When the liquid is discharged, the tip of the gun is attached to the wall of the container, and the angle is maintained at 30-45° with the wall of the container. The pipetting volume is the most accurate.

Maintenance Tips

Regular Cleaning: Keep the pipette clean by following the manufacturer’s cleaning instructions. Regular cleaning can prevent buildup and contamination.
Routine Calibration: Schedule regular calibration to ensure accuracy and precision.
Proper Storage: Store the pipette vertically using a pipette stand to prevent damage to internal components.

Troubleshooting Checklist

1. Inspect the Pipette and Tips:
Check for visible damage to the pipette and tips.
Ensure tips are properly seated and compatible with the pipette.

2. Check Seal and O-rings:
Inspect and replace worn or damaged seals and O-rings.

3. Verify Calibration:
Perform a calibration check and adjust if necessary.

4. Review Pipetting Technique:
Ensure consistent and proper pipetting techniques are being followed.

Hawach can avoid the three major issues of appeal. It combines operational and ergonomic design with advanced and innovative materials and is the perfect manual pipette for scientists who prefer to use the central pipetting button to meet the high operational requirements of the laboratory. All the features required by life science users: Robust construction, one-handed operation, complete sterilization, high precision, and easy calibration technology for long-term reliability.